Do the amino acids have to be pre-activated for spotting?No, the
MultiPep RS can
activate amino acid derivatives specifically for each cycle. There is a
dedicated activation vial for each amino acid. Of course, pre-activated
amino acids can also be used.
Are the membranes deprotected in the instrument?
Yes, the MultiPep RS spots
piperidine solution on the membrane to take the Fmoc protection group
off in each cycle. The membrane can stay in the instrument for the
entire synthesis.
How is the membrane washed and dried for the next coupling?
The
membrane is clamped tightly in a holder which forms a shallow tray
above it. The bottom has tiny holes to remove solvent by vacuum
aspiration. Solvent is delivered on the upper face of the membrane and
pulled through by vacuum. A longer extraction time is being applied to
remove methanol from the final wash and to dry the membrane.
Is the rinsing action sufficient if a full set of peptides is built on the membrane?
Yes,
the flow-through mode of washing is very effective and makes economic
use of the solvents. Actually it is even more effective than washing
manually in plastic trays.
Can we still use Bromophenol blue (BPB) to monitor the reactions?
The MultiPep RS can
be programmed to deliver BPB at the appropriate steps in the cycle to
stain amino groups on the peptides assembled. The stain is removed with
the next coupling reaction as it is done in manual procedure, too.
Is rinsing only applied on the spot where reagents were delivered?
No,
the entire membrane must be washed, even if there is only one single
spot to be coupled. In general, the peptides will be of similar length
and a significant portion of the membrane is still being spotted.
Is there any possibility of piperidine carryover?
As
piperidine is used in a very large excess it must be removed
thoroughly. The washing procedure is designed to remove all traces of
the reagent before the next coupling will be started.
Can the MultiPep be adjusted to make more than 600 spots on each membrane?
So
far, we have worked only with up to 600 spots per membrane which is
relatively easy. Very dense arrays may require manual pre-activation
(once a day) as the very fine needle tip is not suited for the mixing
necessary in the activation procedure.
Is it possible to use a Boc-Lys(Fmoc)-Pro linker and cleave diketopiperazine peptides for QC/QA?
Yes, this can be done and we can supply the respective membrane.
What is the synthesis time for two sets of 600 each 10-mer peptides?
With
a recommended double coupling of 2x30 min, 2x5 min deprotection and all
washing and drying steps, the total cycle time is about 2.5-3 hours and
the two sets of 10mers would be assembled in a little more than a day.
What is the cost of consumables and solvents?
Amino acid derivative sets and membranes are similar to those used in the semi-automated SPOT synthesis on our AutoSpot SL
robot. The amount of piperidine used is considerably less (it accounts
for most of the overall costs!) and the volume of solvents is also a
bit less than in the manual procedure.
Can the synthesis be interrupted and started again?
The
MultiPep RS can
be stopped and re-started at any time. Several convenient options allow
manual procedures (e. g. Bromophenol blue staining or introduction of
very unusual derivatives) and safe re-starting in any cycle.